Logo
User: Guest  Login
Authors:
Eschbaumer, Stephan; Bergmaier, Andreas; Dollinger, Günther 
Document type:
Zeitschriftenartikel / Journal Article 
Title:
A position sensitive time of flight detector for heavy ion ERD 
Collection title:
The 22nd International Conference on Ion Beam Analysis (IBA 2015) 
Journal:
Nuclear Instruments and Methods in Physics Research Section B 
Volume:
371 
Conference title:
International Conference on Ion Beam Analysis (22., 2015, Opatija) 
Venue:
Opatija, Croatia 
Year of conference:
2015 
Date of conference beginning:
14.06.2015 
Date of conference ending:
19.06.2015 
Year:
2016 
Pages from - to:
125-131 
Language:
Englisch 
Keywords:
Elastic recoil detection ; ERDA ; Time of flight detector ; Position sensitive detector ; Electrostatic lens 
Abstract:
Abstract A new 2D position sensitive time of flight detector for heavy ion ERD has been developed. The detector features separate time and position measurement in a straight geometry. An electrostatic lens focuses the secondary electrons ejected from a carbon foil onto a channel plate stack maintaining the position information despite the electron momentum distribution. For position readout a 2D Backgammon anode is used. A position resolution of <0.6 mm (FWHM) and a time resolution of 96 ps (FWHM) is demonstrated. 
ISSN:
0168-583X 
Department:
Fakultät für Luft- und Raumfahrttechnik 
Institute:
LRT 2 - Institut für angewandte Physik und Messtechnik 
Chair:
Dollinger, Günther 
Open Access yes or no?:
Nein / No