Fellinger, Eva; Brandt, Tom; Creutzburg, Justin; Rommerskirchen, Tessa; Schmidt, Annette 
Dokumenttyp:
Zeitschriftenartikel / Journal Article 
Titel:
Analytical Performance of the FreeStyle Libre 2 Glucose Sensor in Healthy Male Adults 
Zeitschrift:
Sensors 
Jahrgang:
24 
Heftnummer:
17 
Jahr:
2024 
Seiten von - bis:
5769 
Sprache:
Englisch 
Stichwörter:
CGM ; glycemic control ; MARD ; oral glucose tolerance test ; sensor accuracy 
Abstract:
Continuous Glucose Monitoring (CGM) not only can be used for glycemic control in chronic diseases (e.g., diabetes), but is increasingly being utilized by individuals and athletes to monitor fluctuations in training and everyday life. However, it is not clear how accurately CGM reflects plasma glucose concentration in a healthy population in the absence of chronic diseases. In an oral glucose tolerance test (OGTT) with forty-four healthy male subjects (25.5 ± 4.5 years), the interstitial fluid glucose (ISFG) concentration obtained by a CGM sensor was compared against finger-prick capillary plasma glucose (CPG) concentration at fasting baseline (T0) and 30 (T30), 60 (T60), 90 (T90), and 120 (T120) min post OGTT to investigate differences in measurement accuracy. The overall mean absolute relative difference (MARD) was 12.9% (95%-CI: 11.8–14.0%). Approximately 100% of the ISFG values were within zones A and B in the Consensus Error Grid, indicating clinical accuracy. A paired t-test revealed statistically significant differences between CPG and ISFG at all time points (T0: 97.3 mg/dL vs. 89.7 mg/dL, T30: 159.9 mg/dL vs. 144.3 mg/dL, T60: 134.8 mg/dL vs. 126.2 mg/dL, T90: 113.7 mg/dL vs. 99.3 mg/dL, and T120: 91.8 mg/dL vs. 82.6 mg/dL; p < 0.001) with medium to large effect sizes (d = 0.57–1.02) and with ISFG systematically under-reporting the reference system CPG. CGM sensors provide a convenient and reliable method for monitoring blood glucose in the everyday lives of healthy adults. Nonetheless, their use in clinical settings wherein implications are drawn from CGM readings should be handled carefully. 
Article-ID:
5769 
Fakultät:
Fakultät für Humanwissenschaften 
Institut:
HUM 5 - Institut für Sportwissenschaft 
Professur:
Schmidt, Annette 
(Forschungs)einrichtung UniBw M:
dtec.bw 
Open Access ja oder nein?:
Ja / Yes 
Art der OA-Lizenz:
CC BY 4.0 
Sonstige Angaben:
Die Veröffentlichung wurde finanziell unterstützt durch die Universität der Bundeswehr München und die DFG. 
Wenn Sie Schwierigkeiten haben, das Dokument zu öffnen, versuchen Sie auch bitte diesen Link